Slot: center_region_intensity_fraction
The fraction of the upper intensity range occupied by the center of illumination region. That is, a value of 0.1 means that the center of illumination occupies the top 10% of the intensity range. microscope-metrics tries to adapt the threshold of this intensity so that the area occupied is not too large. One value per channel.
URI: microscopemetrics_schema:center_region_intensity_fraction Alias: center_region_intensity_fraction
Applicable Classes
Name | Description | Modifies Slot |
---|---|---|
FieldIlluminationKeyMeasurements | no |
Properties
-
Range: Float
-
Multivalued: True
-
Minimum Value: 0
-
Maximum Value: 1
Identifier and Mapping Information
Schema Source
- from schema: https://MontpellierRessourcesImagerie.github.io/microscopemetrics-schema
Mappings
Mapping Type | Mapped Value |
---|---|
self | microscopemetrics_schema:center_region_intensity_fraction |
native | microscopemetrics_schema:center_region_intensity_fraction |
LinkML Source
name: center_region_intensity_fraction
description: The fraction of the upper intensity range occupied by the center of illumination
region. That is, a value of 0.1 means that the center of illumination occupies the
top 10% of the intensity range. microscope-metrics tries to adapt the threshold
of this intensity so that the area occupied is not too large. One value per channel.
from_schema: https://MontpellierRessourcesImagerie.github.io/microscopemetrics-schema
rank: 1000
alias: center_region_intensity_fraction
domain_of:
- FieldIlluminationKeyMeasurements
range: float
multivalued: true
minimum_value: 0.0
maximum_value: 1.0